LABORATORY CONDUCTIVITY / SALINITY METER CC-502


The most important features




Detailed description


Measures conductivity, salinity, resistivity and temperature. Full conductivity measuring range enables measurements in clear water as well as in high concentration samples. The most recent version of this model has been updated and has more possibilities which facilitate usage and increase accuracy.

Characteristic features:




The set includes CT2B-121 temperature probe with Pt-1000B resistor and accurate ECF-1 conductivity cell. Measuring range: 0 ÷ 400 mS/cm is sufficient for conductivity measurements in majority of liquids of maximal concentration, e.g. aqueous soil extracts and water with grease or oil. Metal electrodes are easy to clean. Plastic housing protects from mechanical damage.

In comparison with the CC-505 meter, the CC-502 model is equipped with a smaller display.


TECHNICAL DATA
Function Conductivity Salinity Resistivity Temperature
Range 0 ÷ 1999.9 mS/cm
autorange - 6 subranges
NaCl 0 ÷ 296 g/l
KCl 0 ÷ 239 g/l
0.500Ωcm ÷ 200MΩcm -50.0 ÷ 199.9 °C
Accuracy (± 1 digit) up to 19.999 mS/cm ±0.1 %*,
above 20,00 mS/cm: ±0.25 %*
±2.00 %* ± 2%
of measured value*
±0.1 °C**
Temp. compensation -5 ÷ 70°C -5 ÷ 70°C -5 ÷ 70°C -
α coefficient 0.00 ÷ 10.00 %/°C 0.00 ÷ 10.00 %/°C 0.00 ÷ 10.00 %/°C -
K constant range 0,010 ÷ 19,99 cm-1 0,010 ÷ 19,99 cm-1 0,010 ÷ 19,99 cm-1 -
Printer thermal, width = 60 mm
Power supply 6 V / 2 A power adapter
Dimensions (mm) L = 200, W = 180, H = 20/50
Weight 650 g
PC connection USB

* The accuracy of the meter only.
** The accuracy of the meter only. The total error includes the meters and probe's accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8°C , with Pt-1000A resistor: ±0.35°C.